DATA TRANSMISSION |
 |
Interfaces
|
The modules on this page can be used with the following devices: P-OX100, P-OX100L, PEARL100, PEARL100L, VITRO and CAP10. The measured values are transmitted online (e.g. during measurement). For limitations of a particular device, please see the detailed description below.
|
Available Interfaces
|
|
Ethernet Interface with Patient Isolation
|
|
Interface for Nurse-Call Systems
|
|
Interface for the POX·Net System
|
|
Interface for Analog Recorder with Patient Isolation
|
|
Network Interface
|
With this interface, all devices except for the CAP10 can be integrated into an ethernet system. Each device has his own, unique IP address and can be connected on this address using the TCP/IP or UDP protocol. The interface contains an additional patient isolation. The connection is done over a twisted pair cable to a network switch or -hub.
|
|
|
Nurse·Net Interface
|
This interface connects all devices except for the CAP10 to a nurse-call system. The adapter box contains a relay with a SPDT (singel pole, double throw) contact. Connection is done with a Mini-DIN connector.
|
|
|
POX·Net Interface
|
With this interface, the devices: P-OX 100, P-OX 100L, PEARL100 and PEARL100L can be integrated into the POX·Net system. The POX·Net is a centralized patient monitoring system that works using radio transmission of the patient data. Up to six instruments can be connected to one POX·Net central unit. Several central units can work in parallel without interfering with each other.
|
|
|
Analog Interface
|
This interface module is an isolated three channel analog output. The following channels are available:
P-OX100, PEARL100: Pulse, SpO2, EKG Curve
P-OX100L,PEARL100L: Pulse, SpO2, Plethsymographic waveform
VITRO: Pulse, SpO2, Plethysmographic waveform
CAP10: Respiration rate, etCO2, Capnogram
The recorder is connnected with 4mm banana plugs.
Order number: 08050 Adapter for P-OX100(L) and PEARL100(L) 08060 Adapter for CAP10
|
|
|
|